Nuclear Medicine Phantoms

Standardized Performance Testing for PET & SPECT Systems

NEMA IEC NM Body Phantom

The NEMA IEC NM Body Phantom is designed for accurate, standardized performance testing and calibration of PET and SPECT nuclear medicine imaging systems. It ensures compliance with NEMA and IEC quality assurance standards while supporting reliable validation and verification of imaging performance.

Ideal for routine QA programs, annual scanner evaluations, and multi-system harmonization, the phantom replicates human anatomical features and tissue characteristics to provide realistic and reproducible testing conditions.

SpecificationDetails
Compliance StandardsNEMA NU2, NEMA NU1, IEC 61675-1, IEC 61675-2
DesignHuman torso-sized elliptical cylinder
TargetsMultiple spherical inserts including additional 28 mm sphere
MaterialLung tissue-mimicking material
Software CompatibilityCompatible with SNMMI PAT (Phantom Analysis Toolkit)
Access SystemCam-Lock IEC lid with relief port
Upgrade OptionExpanded testing capabilities via Upgrade Kit

 

Key Advantages

Applications

Suitable for PET and SPECT system acceptance testing, routine quality assurance programs, annual scanner validation, and harmonization across multiple imaging systems in nuclear medicine departments and research facilities.

Graves Phantom

Quantitative SPECT Tissue Characterization

Bar Phantoms

Four-quadrant Bar Phantoms

Flangeless Deluxe PET and SPECT Phantoms

The flangeless SPECT Phantom meets the requirements set by ACR.

Jaszczak SPECT Phantom

Provides consistent performance information for any SPECT or PET system.

PET Sensitivity Phantom™ NEMA

Six Concentric aluminum tubes used to detect camera sensitivity in PET.

PET Scatter Phantom™ NEMA

Determine the imaging systems relative sensitivity to scatter radiation.

Cardiac Insert

Provides a multi-function simulation of the left ventricle, and can be used to evaluate SPECT imaging of cold defects within the “myocardium.”

Triple Line Insert

Used to produce three 1 mm diameter parallel lines of tracer material spaced 7.5 cm apart.

NEMA SPECT Triple Line Source Phantom ™

Designed in accordance with the recommendations by NEMA to standardize the measurement of reconstructed spatial resolution of SPECT.

Neck Phantom

Made of clear Lucite designed to represent a patient’s neck. The phantom has a two part insert that allows counting in a bottle or vial, as well as capsule counting. Meets all suggested requirements for use in counting a thyroid uptake standard source.

CT Performance Phantom

Meets guidelines in AAPM Report #1 for performance and quality assurance of CT scanners.

Specphan Phantom

Simplifies and speeds the acquisition set-up time and increases the reproducibility of the QC measurements.